Journal of Computational Methods in Engineering

Journal of Computational Methods in Engineering

Investigating the Effect of Bidirectional Functionally Graded Structure on the Mechanical Behavior of Bone and Fatigue Life of Dental Implants

Document Type : Original Article

Authors
Department of Mechanical Engineering, Isfahan University of Technology, Isfahan, Iran.
Abstract
Nowadays, due to high quality and efficiency of dental implants, implantation is considered as the best method for replacing the missing teeth. However, some challenges such as the stress-shielding phenomenon can affect the successfulness of this approach. In this research, the effect of a bidirectional functionally graded titanium-hydroxyapatite structure on the mechanical properties and fatigue life of a dental implant screw, by providing a suitable function for changing the mechanical properties from titanium to hydroxyapatite in the implant screw, was investigated using the finite element simulation. The aim of this study is to compare the performance of the bidirectional functionally graded implant screw with unidirectional functionally graded and titanium implant screws, to reduce the stress-shielding phenomenon and evaluate its fatigue life. Some simplifying assumptions such as isotropic linear elastic behavior for the jawbone were employed. The implant screw and the abutment were modeled as an integrated unit. A rigid ceramic crown with simplified geometry was modeled tied to the abutment. For modeling the jawbone geometry, the section of the lower jaw in the molar teeth area was adopted. The influence of adjacent teeth was ignored, and a single tooth was modeled within the jawbone. The results indicate that the bidirectional functionally graded structure for the screw may reduce the stress-shielding effect. Moreover, compared to the unidirectional functionally graded case, the displacements experienced by the jawbone decrease and hence, prevent damaging the adjacent teeth and jawbone. It is worth mentioning that the proposed bidirectional functionally graded screw satisfy the infinite fatigue life condition, as it should be.
Keywords
Subjects

1.
·         Brånemark, P.I., Breine, U., Adell, R., Hansson, B., Lindström, J. and Ohlsson, Å. Intra-osseous Anchorage of Dental Prostheses: I. Experimental Studies. Scandinavian Journal of Plastic and Reconstructive Surgery. 1969; 3: 81-100.
2.
Bayata, F. and Yildiz, C., The effects of design parameters on mechanical failure of Ti-6Al-4V implants using finite element analysis. Engineering Failure Analysis. 2020; 110: 104445.
3.
Jafari, B., Katoozian, H.R., Tahani, M. and Ashjaee, N., A Comparative Study of Bone Remodeling Around Hydroxyapatite-coated and Novel Radial Functionally Graded Dental Implants Using Finite Element Simulation. Medical Engineering and Physics. 2022; 102: 103775.
4.
Ashwinkumar, A. Review on Functionally Graded Materials and Various Theories. International Research Journal of Engineering and Technology (IRJET). 2017; 4(9): 890-893.
5.
Watari, F., Yokoyama, A., Saso, F., Uo, M. and Kawasaki, T. Fabrication and Properties of Functionally Graded Dental Implant. Composites Part B: Engineering. 1997; 28(1-2): 5-11.
6.
Shirazi, H.A., Ayatollahi, M., Karimi, A. and Navidbakhsh, M. A Comparative Finite Element Analysis of Two Types of Axial and Radial Functionally Graded Dental Implants with Titanium One Around Implant-bone Interface. Science and Engineering of Composite Materials. 2017; 24: 747-754. https://doi.org/10.1515/secm-2015-0392
7.
Taheri, R.A., Jarrahi, A., Farnoosh, G. and Karimi, A. A Comparative Finite Element Simulation of Stress in Dental Implant–bone Interface Using Isotropic and Orthotropic Material Models in Three Mastication Cycles. Journal of the Brazilian Society of Mechanical Sciences and Engineering. 2018; 40: 1-12. https://doi.org/10.1007/s40430-018-1409-9
8.
Ouldyerou, A., Merdji, A., Aminallah, L., Mehboob, H., Mehboob, A., Roy, S., Goswami, T., Mukdadi, O.M. and Tarlochan, F. Functionally Graded Ceramics (FGC) Dental Abutment with Implant-supported Cantilever Crown: Finite Element Analysis. Composites Communications. 2023; 38: 10-15. https://doi.org/10.1016/j.coco.2023.101514
9.
Elleuch, S., Jrad, H., Wali, M. and Dammak, F. Mandibular Bone Remodeling Around Osseointegrated Functionally Graded Biomaterial Implant Using Three Dimensional Finite Element Model. International Journal for Numerical Methods in Biomedical Engineering. 2023; 39(9): 37-50.
https://doi.org/10.1002/cnm.3750
10.
Moayedee, Y., Nikzad, L. and Majidian, H. Exploration into the Microstructural, Mechanical, and Biological Characteristics of the Functionally Graded 3Y-TZP/Ti6Al4V System as a Potential Material for Dental Implants. Journal of the Mechanical Behavior of Biomedical Materials. 2024; 151: 106380.
11.
Dabaja, R., Sayeed Khan, A. and Banu, M. Mechanical Behavior Analysis of a Functionally Graded Porous Dental Implant. International Manufacturing Science and Engineering Conference, American Society of Mechanical Engineers. 2024;  88100.
12.
Rajaeirad, M., Fakharifar, A., Posti, M.H., Khorsandi, M., Watts, D.C., Elraggal, A., Ouldyerou, A., Merdji, A. and Roy, S. Evaluating the Effect of Functionally Graded Materials on Bone Remodeling Around Dental Implants. Dental Materials. 2024; 40(5): 858-868.
13.
Ouldyerou, A., Mehboob, H., Merdji, A., Aminallah, L., Mehboob, A. and Mukdadi, O.M. Biomechanical Analysis of Printable Functionally Graded Material (FGM) Dental Implants for Different Bone Densities. Computers in Biology and Medicine. 2022; 150:106-111.
14.
Akbas, O., Greuling, A. and Stiesch, M. The Effects of Different Grading Approaches in Additively Manufactured Dental Implants on Peri-implant Bone Stress: A Finite Element Analysis. Journal of the Mechanical Behavior of Biomedical Materials. 2024;  154: 106530.
15.
Niroomand, M.R., Arabbeiki, M. and Rouhi, G. Optimization of Thread Configuration in Dental Implants Through Regulating the Mechanical Stimuli in Neighboring Bone. Computer Methods and Programs in Biomedicine. 2023; 231: 107376.
 https://doi.org/10.1016/j.cmpb.2023.107376
16.
Rizzo, P. A Review on the Latest Advancements in the non-Invasive Evaluation/Monitoring of Dental and Trans-femoral Implants. Biomedical Engineering Letters. 2020; 10: 83-102.
https://doi.org/10.1007/s13534-019-00126-8
17.
Kasani, R., Attili, B.K.R.S., Dommeti, V.K., Merdji, A., Biswas, J.K. and Roy, S. Stress Distribution of Overdenture Using Odd Number Implants–A Finite Element Study. Journal of the Mechanical Behavior of Biomedical Materials. 2019; 98: 369-382. https://doi.org/10.1016/j.jmbbm.2019.06.030
18.
Mehboob, H., Tarlochan, F.,Mehboob, A. and Chang, S.H. Finite Element Modelling and Characterization of 3D Cellular Microstructures for the Design of a Cementless Biomimetic Porous Hip Stem. Materials and Design. 2018; 149: 101-112.
19.
Yang, Y., Chang, E. and Lee, S. Mechanical Properties and Young's Modulus of Plasma‐Sprayed Hydroxyapatite Coating on Ti Substrate in Simulated Body Fluid. Journal of Biomedical Materials Research Part A: An Official Journal of The Society for Biomaterials. The Japanese Society for Biomaterials, and The Australian Society for Biomaterials and the Korean Society for Biomaterials. 2003; 67(3): 886-899.
20.
UNE, E., 14801: 2017; Dentistry—Implants—Dynamic Loading Test for Endosseous Dental Implants (ISO 14801: 2016). International Organization for Standardization. Plzen, Czech Republic, 2016.
21.
Baufeld, B. and Van der Biest, O. Mechanical Properties of Ti-6Al-4V Specimens Produced by Shaped Metal Deposition. Science and Technology of Advanced Materials. 2009; 10(1): 015008.
22.
Torabinejad, B., Mohammadi Rovshandeh, J., Davachi, S.M. and Zamanian, A. Synthesis and Characterization of Nanocomposite Scaffolds Based on Triblock Copolymer of L-lactide, ε-caprolactone and Nano-hydroxyapatite for Bone Tissue Engineering. Materials Science and Engineering. 2014; 42: 199-210.
23.
Sun, F., Lv, L.T., Cheng, W., Zhang, J.L., Ba, D.C. and Song, G.Q. Effect of Loading Angles and Implant Lengths on the Static and Fatigue Fractures of Dental Implants. Materials. 2021; 14: 5542.
https://doi.org/10.3390/ma14195542
24.
Janeček, M., Nový, F., Harcuba, P., Stráský, J., Trško, L., Mhaede, M. and Wagner, L. The Very High Cycle Fatigue Behaviour of Ti-6Al-4V Alloy. Acta Physica Polonica A. 2015; 128(4): 497-502.
 